Maintenance software for laboratories
Updated
FleetZap schedules service for every lab instrument, records readings and signatures on each check and keeps a full history you can pull up when an auditor or customer asks.
Ultra-low freezers that drift warm
Laboratories depend on freezers, refrigerators, incubators, fume hoods, biosafety cabinets, centrifuges, balances, autoclaves and analytical instruments, plus the building systems behind them. Most run on calendar service intervals. Your own staff handle some of the work and outside vendors handle much of the rest, and many checks mean reading a temperature, an airflow or a weight against a limit.
Your staff read an ultra-low freezer’s display each morning and write the number on a sheet taped to its door. Nothing compares that number to a limit, so a freezer that runs a little warmer each day can go a week before anyone spots the trend. Vendor contracts and visit dates sit with different people, and one instrument’s history ends up split between binders and emails.
Ultra-low freezers fail in predictable ways. Frost builds up on the door gaskets and inner doors and lets warm air in, a clogged condenser filter makes the compressors work harder, and the alarm’s backup battery dies without anyone testing it. Manufacturers ask for condenser filter cleaning and gasket care on a short calendar interval, with a defrost when the frost gets thick. Biosafety cabinets are a different case: a qualified certifier, usually an outside vendor, certifies them in the field on a schedule your biosafety officer sets. Both fit a date-based schedule with a procedure, and the certifier’s visit can be a vendor work order with its cost.
Spreadsheet intervals, vendor visits and drifting readings
- Many instruments, many intervals: Freezers, hoods and balances each have their own interval, and the master list lives in a spreadsheet one person maintains.
- Outside service vendors: Vendors do much of the work, so visit dates and invoices sit in someone’s inbox and not on the instrument’s record.
- Readings against limits: Temperatures, airflow and weights get logged, but nobody flags a value that drifts out of range.
- Showing the history: A reviewer asks about one centrifuge, and its records are split between a binder and the vendor’s emails.
Instrument PM, readings and vendor records
- Calendar PM for every instrument: Schedule service by date for one instrument, a whole category or a list of units, with a procedure attached.
- Readings with limits: Reading steps have a unit and a min/max; an out-of-range value fails the step and opens a repair work order.
- Vendor work with cost: Work orders can go to an outside vendor with its cost and reference, and vendor turnaround and spend are tracked.
- Audit log and signatures: Signature steps record who signed off, and the audit log records every change: who, when and which fields.
- Locations by building and room: Nested locations place each instrument in its building, floor and room.
- Requests from lab staff: Staff scan the QR label on the instrument and report the problem with photos. They don’t need a FleetZap account to do it.
Freezers, hoods, centrifuges and balances
- Ultra-low freezers
- Refrigerators and incubators
- Fume hoods
- Biosafety cabinets
- Centrifuges
- Balances
- Autoclaves
- Analytical instruments
Overdue service and lab vendor turnaround
- PM compliance: Check compliance by instrument category; a hood or biosafety cabinet with lapsed certification may need tagging out.
- Overdue service: Overdue vendor visits often sit in someone’s inbox; chase the vendor and log the new date on the instrument.
- Vendor turnaround: Compare turnaround across vendors for the same instrument type before you renew a service contract.
- Availability: Low availability on a shared centrifuge or freezer hits several groups at once; consider a backup unit for those.
Free tools for this
- Equipment lease calculator: Estimated lease payment, and the total cost of leasing against buying with a loan
- Equipment financing and loan calculator: Monthly payment, total interest and total cost of an equipment loan
- OEE calculator: Overall equipment effectiveness from availability, performance and quality
- MTBF calculator: Mean time between failures, mean time to repair and availability
FleetZap Maintenance
Work orders, PM, parts, compliance, assets and KPIs. Explore FleetZap Maintenance.
Frequently asked questions
Is FleetZap validated for regulated labs?
FleetZap doesn’t claim compliance with any specific regulation and isn’t sold as a validated system. It provides an audit log, signature steps on procedures and a full history for each instrument; validating it for your quality system is your responsibility.
Can we track service done by outside vendors?
Yes. A work order can be assigned to an outside vendor with its cost and a reference number. That cost adds to the instrument’s maintenance cost, and the Vendors page shows each vendor’s jobs, spend and turnaround, which helps when a service contract comes up for renewal.
Can technicians record temperatures and other readings?
Yes. Reading steps have a unit and an allowed range, so a freezer check can ask for the cabinet temperature with its limits. An out-of-range reading fails the step and opens a repair work order linked to the check, and the reading stays on the instrument’s history.
Can lab staff report a broken instrument?
Yes. Put a QR label on each instrument. Anyone in the lab can scan it, describe the problem and add photos without a FleetZap account. Maintenance reviews the request and approves it into a work order, which already knows which instrument and room it belongs to.